Single Beam UV Visible Spectrophotometer
Single Beam UV Visible Spectrophotometer
Wide spectral coverage and high resolution is achieved by using a powerful light source and a grating-based monochromator.
A conveniently placed control enables the user to quickly orient the grating to provide monochromatic light of desired wavelength.
The electronic processing circuits and power supplies are solid-state / LSI-based to ensure high accuracy, stability and long life.
104
Specification:
- Wavelength
- Range – 340 to 960 rim
- Wavelength
- Accuracy – ±5 nm
- Wavelength
- Readout – 5 nm
- Wavelength Display – Dial
- Spectral Bandwidth – 10 nm
- Photometric Accuracy – +0,005 Abs at 1.000 Abs
- Photometric Resolution – 0.1%T, 0.001 Abs
- Photometric Readout – 4 digit Seven Segment LED
Grating – 600 linesimm - Source – Tungsten halogen lamp
- Detector – Wide Range Silicon Photodiode
- Sample Holder – Single position 10 mm cuvette holder
- Min. Sample Volume – 1 ml in 4 ml cuvette
- Measuring Mode -Auto
a)%Transmittance
b) Absorbance
c) Concentration 230V, - Power – +10%, 50Hz 420(W) X 310(D)
- Dimension – X 125(H) mm
- Weight – 6.5 Kg (Approx)
- Accessories – Two Glass Cuvettes
- Optional Accessories -Single Position 20150 mm Cuvette Holder (Factory Installed)
105
Specification:
- Wavelength
- Range – 340 to 960 rim
- Wavelength
- Accuracy – ±5 nm
- Wavelength
- Readout – 5 nm
- Wavelength Display – Dial
- Spectral Bandwidth – 10 nm
- Photometric Accuracy – +0,005 Abs at 1.000 Abs
- Photometric Resolution – 0.1%T, 0.001 Abs
- Photometric Readout – 4 digit Seven Segment LED
Grating – 600 linesimm - Source – Tungsten halogen lamp
- Detector – Wide Range Silicon Photodiode
- Sample Holder – Single position 10 mm cuvette holder
- Min. Sample Volume – 1 ml in 4 ml cuvette
- Measuring Mode -Manual
a)%Transmittance
b) Absorbance
c) Concentration 230V, - Power – +10%, 50Hz 420(W) X 310(D)
- Dimension – X 125(H) mm
- Weight – 6.5 Kg (Approx)
- Accessories – Two Glass Cuvettes
- Optional Accessories -Single Position 20150 mm Cuvette Holder (Factory Installed)
106
Specification:
- Wavelength – 340 to 960 nm
Absorbance Range – -0.3 to 2.500 Abs
Spectral Bandwidth – 5nm
Wavelength Resolution – 5nm
Wavelength Display – Dial
Wavelength Accuracy – +5 nm
Photometric Resolution – 0.1%T
Photometric Accuracy – ±0.005 Abs at lAbs
Modes – %T, Abs, Concentration & Factor
Concentration – 0.000 to9999 with floating Decimal point
K factor – 0.000 to9999 with floating Decimal point
Photometric Readout – 4 Digit LED Display
0 & 100%T Setting – Manual
Grating – 600 linesirnm
Detector – Si-Photo- Diode
Source – Halogen Lamp
Sample holder – l Omm path length,
Two Position Manual Sample Changer
Power – 230V ±-10%, 50Hz
Dimension Weight – 420(W) X310 (D) X 125(H) mm 6.5 Kg (Approx)
Accessories – Two Glass Cuvettes
166
Specification:
- Wavelength – 340 to 960 nm
Absorbance Range – -0.3 to 2.500 Abs
Spectral Bandwidth – 5nm
Wavelength Resolution – 5nm
Wavelength Display – 3 Digital dial
Wavelength Accuracy – +2 nm
Photometric Resolution – 0.1%T
Photometric Accuracy – ±0.005 Abs at lAbs
Modes – %T, Abs, Concentration & Factor
Concentration – 0.000 to9999 with floating Decimal point
K factor – 0.000 to9999 with floating Decimal point
Photometric Readout – 4 Digit LED Display
0 & 100%T Setting – Manual
Grating – 600 linesirnm
Detector – Si-Photo- Diode
Source – Halogen Lamp
Sample holder – l Omm path length,
Two Position Manual Sample Changer
Power – 230V ±-10%, 50Hz
Dimension Weight – 420(W) X310 (D) X 125(H) mm 6.5 Kg (Approx)
Accessories – Two Glass Cuvettes
167
Specification:
- WAVELENGTH
Range-340 to 999 nm
Accuracy-±2.0 nm
Readout-1 nm step
Bandwidth-5.0 nm
PHOTOMETRIC
Range–0.04 to 2.5 Abs
Accuracy-10.005 Abs at 1.0 Abs
Repeatability -±0.005 Abs
Resolution-0.1 in %T mode, 0.001 Abs in Abs mode
Digital Readout-2 line 20 characters LCD
Drift- ±0.003 Abs/hour at 500 nm ±0.005 Abs
Baseline Flatness Stray Light -Less than 0.3%T at 370 rim
GRATING – 600 lines/mm
SOURCE – Tungsten-Halogen lamp
DETECTOR – Silicon Photodiode
SAMPLE HOLDER-Two-position manual sample changer or factory installed single position holder for
20 mm / 50 mm (Max) path length cell.
MINIMUM. VOLUME
0,100%T, ORDER-1 ml in 4 ml cuvette
CUTOFF SETTING – automatic
OPERATING MODES Automatic
i) Single Wavelength
ii) Multi Wavelength (max 6)
iii) Scan
MEASURING MODES-i) Absorbance
ii) %Transmittance
iii) Concentration (with factor and curve fitting)
K-Factor Range Concentration Range – 0.100 to 9999 with floating Decimal point
0.100 to 9999 with floating Decimal point
DATA PROCESSING- 1) Peak Pick (6 peaks in Abs value)
2) Spectrum printout
(on Epson compatible Dot Matrix printer)
DATA PRESENTATION
Display-2 line 20 char. Alphanumeric LCD
Hard Copy-On printer (Graph and Data both)
PRINTER PORT-Epson compatible 80 Column Dot Matrix
RS 232 PORT -PC Link data transfer
POWER-230V 110%, 50 Hz
DIMENSIONS-420 (W) X 310 (ID) X 125 (H) mm
WEIGHT-6.5 kg (Approx)
ACCESSORIES-2 matched 10 mm path length Glass cuvettes
169
Specification:
- Wavelength – 320-990nm
Absorbance Range – -0.3 to 2.500 Abs.
Spectral Bandwidth – 5nm
Wavelength Resolution – 1nm
Wavelength Display – Digital
Wavelength Accuracy – ±2nm
Photometric Resolution – 0 . 1`)/0 T
Photometric Accuracy at 1 Abs. – ±0.005 Abs.
Modes – %T, Abs., Concentration & Factor
Concentration – 0.000 to 9999 with floating Decimal point
K factor – 0.000 to 9999 with floating Decimal point
Photometric Readout – 4 Digit LED Display
0& OCr/oT Setting – Auto
Grating – 600 lines/mm
Detector – Si-Photo- Diode
Source – Halogen Lamp 12V-50W
Sample holder – lOmm path length, Two Position Manual Sample Changer
Power (SMPS) – 230V±10°/050Hz
Printer – Printer Port for 80 Column DMP
Optional – 20mm/50mm Single Position Factory Installed Holder For Computer attachment RS232 interface port
1203
Specification:
- WAVELENGTH
Range – 340 to 999 nm
Accuracy – ±2 nm
Readout – 1 nm per step
Bandwidth-5 nm
PHOTOMETRIC
Range-±2.5 Abs
Accuracy-±0.005 Abs at 1.0 Abs
(with NIST filter)
Repeatability Resolution- ±0.005 Abs
Display-0.1%T, 0.001Abs
Drift-4 Line 20 Ch. LCD
Baseline Flatness : ±0.003 Abs / hour at 500 nm
Scan Speed – ±0.005 Abs
Grating
Source – Slow, Medium & Fast
Detector – 600 lines/mm
Sample Holder : Tungsten halogen lamp 2-Silicon Photodiode
Minimum Volume : One (10 mm) fixed for Reference & Sample cuvette holder
Operating Mode :1 ml in 4 ml cuvette with cell riser Single wavelength, Multi wavelength (Max 8), Spectrum scan and Time scan
Measuring Mode :
K-Factor Range Conc .Range– Absorbance, “%Transmittance and Concentration (with by factor and by standard curve fitting)
Data Processing :0.100 to 9999 with floating
decimal point
0.100 to 9999 with floating
decimal point
Stray Light-Peak pick/Valley picks (6 points) Point picks
Printer Port -(8 point), First derivative, Data saving facility (Four user defined files for each mode) Test saving facility (Four user defined files for each mode)
RS232 Port Power -< 0.3%T at 370 nm
Dimension Weight defined files for each mode)
Accessories Optional -Epson compatible 80 column Dot Matrix
Accessories– PC Link data transfer
230 V ±10 % , 50 Hz
450(W) X 360(D) X 130(H) mm
9 Kg (Approx)
Two 10 mm glass cuvette
n PC link software
• Epson compatible D. M. Printer
• Multi cuvette holders for 20/40/50 mm rectangular cuvette
(Factory Installed)
117
Specification:
- OPTICS : Modified Czerny – Turner Monachromator geomety for better aberration correction with Holographic Grating of 1200 lines/mm blazed at 250 nm.
- WAVELENGTH
Range : 200 to 1100 nm
Resolution : 0.1 nm
Accuracy : 1 nm
Repeatability : t 0.5 nm
Bandwidth : 2 nm - PHOTOMETRIC
Range : 0 to 2.5Abs
Accuracy : 0.005 Abs at IA
Repeatability : 0.002A bs at I A
Drift : 0.003 Abs /hr (After 1 hr warmup at 500 nm) - STRAY LIGHT
Less then O. 1% T at 220 nm 8: 370 nm - BASELINE CORRECTION Automatic baseline correction
- BASELINE FLATNESS
0.003 Abs /hr (After 40 minutes warm-ups) - MEASURING MODES
%Transmittance
Absorbance
Concentration by K-factor and Multi-standards - OPERATING MODES
Single Wavelength
Multi Wavelength
Spectrum Scan
Time Scan - DATA PROCESSING : Stand Alone • Peak-Pick • Spectrum printout
- SCAN SPEED : Low, Medium and High
- DATA INTERVAL : Depends on wavelength scan-span & Scan-speed
- SAMPLE HOLDER : 5 position for 10 mm cuvette (automatically) SAMPLE VOLUME 2 ml in 4 ml cuvette or 500 pi in 1 ml cuvette SOURCE : Tungsten Halogen Lamp (320 to 1100 nm)
Deuterium Lamp (200 to 340 nm) - DETECTOR : UV – enhanced wide range solid state photodiode DISPLAY : 20 character, 41ine backlit alphanumeric LCD KEYBOARD :18 keys (11 numeric and 7 functional)
- PRINTER PORT : Epson compatible 80 Column Dot Matrix RS 232 PORT : PC Link data transfer
- POWER 230V ±10%, 50 Hz
- DIMENSIONS : 560 (W) X 425 (D) X 205 (H) mm
- WEIGHT :21 kg (Approx)
- ACCESSORIES : Two matched 10 mm path length Quartz cuvettes
- OPTIONAL ACCESSORIES
Holder for 50 mm or 100 mm cylindrical and rectangular cuvettes
Personal Computer / Laptop
Epson compatible 80 column D.M. Printer
Standard RS 232 interface for Window based PC link Software for Spectrum Display, Storage/Retrieve, Peak-Pick, Point – Pick, Expansion of spectra, I 3r’l& 4’1′ Derivatives, Averaging of Scans,
Subtraction of Two Scans